    Light Conversion-HARPIA-TA 超快瞬態(tài)吸收 / 反射光譜儀


    • 330 nm – 24 μm spectral range
    • Probe delay ranges from 2 ns to 8 ns
    • Pump pulse energies down to nJ
    • Cryostat and peristaltic pump support





    導(dǎo)出PDF

    The HARPIA-TA ultrafast transient absorption spectrometer provides pump-probe measurement functionality in a HARPIA system. Several probe light configurations and detection options are available: from a photodiode for single-wavelength detection to white-light supercontinuum probing combined with spectrally-resolved broadband detection. HARPIA-TA features automation options providing pump and probe beam position tracking and alignment, pump polarization control, supercontinuum generator switching, sample positioning, as well as switching between transient absorption and transient reflection measurements. Broadband probe options cover a 330 – 16000 nm range, while a monochromatic probe can be used up to 24 μm. The probe delay stage is configurable from 2 ns to 8 ns.

    HARPIA-TA features the sensitivity of 0.05 mOD (10-4 ΔT/T) and can be operated at high repetition rates of up to 1 MHz when used with a PHAROS or CARBIDE laser and an ORPHEUS series OPA, which allows the study of transient absorption dynamics with excitation pulse energies down to several nanojoules.

    HARPIA-TA is compatible with cryostats and peristaltic pumps, and the capabilities of the spectrometer are extendable using expansion modules.



    Model

    HARPIA-TA

    Configuration

    UV / VIS / NIR / SWIR

    MIR

    Probe excitation wavelength

    1030 nm

    515 nm

    800 nm

    n/a 1)

    Probe spectral range

    460 – 1600 nm

    350 – 750 nm

    330 – 1400 nm

    190 nm – 16000 nm 2)

    Detection spectral range

    200 – 1100 nm, 900 – 1700 nm, or 900 – 2600 nm

    2 – 13 μm 3)

    Delay range

    2 ns, 4 ns, or 8 ns

    Delay resolution

    2.1 fs, 4.2 fs, or 8.3 fs

    Laser repetition rate

    1 – 1000 kHz

    Temporal resolution

    < 1.4× pump or probe pulse duration, whichever is longer

    Maximum data acquisition rate

    4000 spectra/s

    SNR 4)

    250:01:00


    • A wavelength-tunable source is used instead of a laser-excited white-light continuum.
    • An extended tuning range of ORPHEUS‑HP; see specification for more details. Also applicable to UV/VIS/NIR/SWIR configuration.
      Up to 24 μm available upon request.
      Estimated as the standard deviation of a set of 2500 spectra measured in SCHOTT OG530 glass with 54 nJ, 370 nm pump and > 4.5 mOD at a maximum of the spectrum. Not applicable to all samples and configurations.
    • Up to 24 μm available upon reques.
    • Estimated as the standard deviation of a set of 2500 spectra measured in SCHOTT OG530 glass with 54 nJ, 370 nm pump and > 4.5 mOD at a maximum of the spectrum. Not applicable to all samples and configurations.